Watlow Series 935A User’s Manual 55
935A Software Map
Operations Menu
Prompt Range Default Hidden if * Your Settings
*NOTE:
All prompts, except
[``SP] and [`tAG],
are hidden if [`Ot1] =
[`FPL].
[`tA9] = [`O``];
or if
[`Ot2] [thM] or
[tMs]
[trAc][trAc] [``rL]
[``rh]
Choose an idle Set
Point. Trac is an Idle Set
Point equal to the
Primary Set Point; or
select a value between
Input Type Range Low
and Input Type Range
High.
[idLE]
Idle Set Point Type
[`tA9] = [`O``];
or if
[`Ot2] [thM] or
[tMs]
0
[0000]
Select a countdown time
value, adjustable
between 00:00 and
99:59 hrs:mins or
min:sec.
[tMr]
Countdown Timer
[`tA9] = [`O``];
or if
[`Ot1] and [`Ot2]
[ALM]
Type J Range High[`RL0]
Select a value,
adjustable between
Alarm Low and Range
High.
[`Ah1]
Alarm High
[`tA9] = [`O``];
or if
[`Ot1] and [`Ot2]
[ALM]
Type J Range Low[`Ah1]
Select a value,
adjustable between
Range Low and Alarm
High.
[`ALO]
Alarm Low
[`tA9] = [```A]no[``no] [`YES]
Choose YES to Auto-
tune the 935A.
[`Aut]
Auto-tune
Appears always
75°F or 23°C[``rL] [``rh]
Select a value,
adjustable between Input
Type Range Low and
Input Type Range High.
[``SP]
Primary Set Point