Watlow Series 988LF Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 64
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
2.8 WATLOW Series 988LF User’s Manual
Installation and Wiring, Chapter 2
Figure 2.8d
0-20mA or 4-20mA Process
Universal signal conditioner
98 _ L - _ 2 _ _ - A A _ _
Input impedance: 7
Remote Set Point Option
Figure 2.8c
0-5V
ÎÎ
, 1-5V
ÎÎ
or 0-10V
ÎÎ
(dc) Process
Universal signal conditioner
98 _ L - _ 2 _ _ - A A _ _
Input impedance: 10K
Remote Set Point Option
Figure 2.8b RTD (2 or 3 Wire) 100
Universal signal conditioner
98 _ L - _ 2 _ _ - A A _ _
Remote Set Point Option
20
19
-
+
Figure 2.8a
Thermocouple or 0-50mV (high impedance)
Universal signal conditioner
98 _ L - _ 2 _ _ - A A _ _
Input impedance: 20M
Remote Set Point Option
19
20
+
-
Jumper
#19 to #20
for 2 wire
RTD
19
20
18
S2
S1
DIP Switch
Setting
R, S, B
DIP Settings
J, K, T, N, C, E, D, Pt2,
0-50mV DIP Settings
DIP Switch
Setting
19
20
18
S2
S1
S3
O
N
123
O
N
123
O
N
123
O
N
123
20
18
-
+
DIP Switch
Setting
O
N
123
19
20
+
-
0-50mV
NOTE:
Successful installa-
tion requires five
steps:
Model number and
software choice
(Appendix);
DIP switch set-
tings (Chapter 1);
Sensor match
(Chapter 2 and
Appendix);
Sensor installation
(Chapter 2); and
Wiring (Chapter 2).
Input 2 Wiring
ç
CAUTION:
An external resistor
is required for
0-20mA and 4-20mA
process wiring to
prevent a high in-
rush current which
could burn out the
controllers 7-ohm
resistor. See page
2.4 for recommen-
dations.
Vista de pagina 17